Inhouse product
The following data describe a laboratory water system that produces ultrapure water from standard tap water. Model MSLWP‑F3‑VF10SWISS is the 10 litre per hour variant of the VF series. The specifications are measured under the manufacturer‘s standard test conditions.
Model identification
Model number: MSLWP‑F3‑VF10SWISS
Type: Ultrapure water purification system with pyrogen, RNase, and DNase removal
Production and dispensing
Water‑making speed: 10 litres per hour (at 25°C)
Water flow speed: 1.5‑2 litres per minute (requires pressure bucket)
Water outlets: 2 outlets (RO pure water and Ultrapure water)
Inlet water requirements
Source: Urban tap water
Feed water TDS: <200 ppm
Inlet water pressure: 0.1‑0.4 MPa
Water temperature: 5‑45°C
RO pure water quality (first stage)
Desalination rate: 95‑99% (with new RO module)
Conductivity: ≤10 μS/cm
Ultrapure water quality (final output)
Resistivity (25°C): 18.25 MΩ·cm
Conductivity: ≤0.055 μS/cm (calculated from resistivity)
TOC: <3 ppb
Particles (>0.22 μm): <1 per ml
Microorganism: <1 cfu per ml
Pyrogen: <0.001 EU per ml
RNases: <1 pg per ml
DNases: <5 pg per ml
Physical dimensions
Dimensions (W × D × H): 470 × 450 × 550 mm
Package dimensions (W × D × H): 640 × 550 × 700 mm
Gross weight: 37 kg
Electrical
Power supply: 220V / 50Hz
Power consumption: 120W
Low‑voltage 24VDC main power supply for safety in humid environments
Standard configuration
Main unit (including 1 set of filters)
Built‑in 8L pressure tank with anti‑bacteria transparent inner pouch (replaceable)
End point filter (0.22 μm terminal filter)
Accessory package
Configuration code: ①②③④⑤⑥⑦⑧⑨⑩ (VF series full configuration)

Key elements include:
Type U Korean quick‑insert filters with imported filtering materials
Upgraded self‑cleaning filter with built‑in 10″ 10μm stainless steel filter mesh and silicone soft scraping (no PP cotton filter replacement needed)
Long‑acting KDF filter (10 times the adsorption life of activated carbon)
Imported UF ultrafiltration column for pyrogen and RNA removal
Built‑in feed water pressure test gauge
Three‑way water quality sensor (feed water, RO, UP) with five‑way signal integration
Timing quantitative water taking function
USB Host interface for data export of historical alarm records and water intake records
RO and UP water circulation system with UV sterilisation
Safety features
Over‑temperature audible and visual alarm (integrated)
Low‑voltage 24VDC main power supply for personal safety in wet lab environments
Power‑off parameter memory
Automatic reminder for filter replacement
Scope of application (VF series)
High‑sensitivity ICP/MS
PCR and qPCR
ppt‑level analysis
Isotope analysis
Molecular biology
Life sciences
Disease control centers
Drug testing institutes
Quality inspection institutes
University research laboratories
Various high‑end precision instruments
The specifications above produce measurable improvements in daily laboratory operations. Each point below connects a feature to a practical outcome.
Ultrapure water at 18.25 MΩ·cm for sensitive assays – The two‑stage purification (RO followed by ion exchange, UV, and UF) delivers water at the theoretical maximum resistivity. HPLC mobile phases prepared from this water produce clean, stable baselines. ICP‑MS blanks show no detectable contamination. LC‑MS runs exhibit minimal background noise.
Removal of biological contaminants for molecular work – The VF series removes pyrogen to <0.001 EU per ml, RNases to <1 pg per ml, and DNases to <5 pg per ml. PCR master mixes prepared with this water show no unexpected nuclease degradation. Cell culture media support consistent growth across multiple passages. RNA extraction yields intact, high‑quality nucleic acids.
Self‑cleaning front filter eliminates PP cotton replacement – The built‑in 10μm stainless steel filter mesh with silicone soft scraping cleans itself. The operator turns a knob, and the scraper removes accumulated particles. No more buying and replacing PP cotton filters every few weeks. A laboratory processing 50 litres per day saves approximately 20 filter changes per year.
Long‑acting KDF filter lasts 10 times longer than activated carbon – The KDF (Kinetic Degradation Fluxion) filter removes chlorine and heavy metals while inhibiting bacterial growth. Its 10‑fold longer lifetime means fewer changeouts. The system alerts the user only when the KDF filter actually needs attention, not on a fixed schedule.
Transparent, replaceable inner pressure tank pouch shows water level – The built‑in 8 litre pressure tank has a transparent inner pouch. The operator sees the remaining water volume at a glance. The pouch material is antibacterial, and the pouch itself can be replaced without replacing the entire tank. This avoids the hidden biofilm buildup common in opaque pressure tanks.
USB data export for audit‑ready records – The system stores historical alarm records and water intake records. A technician plugs a USB drive into the host interface and exports the log. Accredited laboratories use this data for internal audits and external inspections. No manual logging, no missing entries.
Pain point: RNase and DNase contamination ruins PCR experiments.
Standard ultrapure water lacks nuclease removal. A single contaminated batch degrades RNA or DNA in sensitive assays. The MASTER SCIENCE MSLWP‑F3‑VF10SWISS ultrapure water purification system (a VF series model with an imported UF ultrafiltration column) removes RNases to <1 pg per ml and DNases to <5 pg per ml. PCR master mixes prepared from this water show intact nucleic acids after thermal cycling. No unexplained band smearing, no degraded controls.
Pain point: Pyrogen contamination causes false results in cell‑based assays.
Endotoxins from Gram‑negative bacteria activate immune cells and alter assay outcomes. The same MASTER SCIENCE MSLWP‑F3‑VF10SWISS ultrapure water purification system (a pyrogen‑removing model with <0.001 EU/ml endotoxin) reduces pyrogen to undetectable levels. Cell culture media and buffer solutions support sensitive cell lines without unexpected activation. Toll‑like receptor assays and cytokine measurements remain accurate.
Optimum Solution is a leading supplier of laboratory equipment, analytical instruments, and process control solutions in Bangladesh. Since our inception, we have built a strong presence in the industry, serving research labs, universities, pharmaceutical companies, and quality control laboratories nationwide. Through continuous dedication to improving our products and services, Optimum Solution has earned the trust and loyalty of our valued clients.